- Research
- Open access
- Published:
Motion resistance in peripheral oxygen saturation monitoring using Biolight Analog SpO2 compared to Masimo SpO2: a non-inferiority study
BMC Anesthesiology volume 24, Article number: 430 (2024)
Abstract
Background
Pulse oximeters are vital for assessing blood oxygen levels but can produce inaccurate readings during patient motion, leading to false alarms and alarm fatigue. Analog SpO2 Technology, which uses analog waveforms to filter motion artifacts, may improve accuracy compared to digital sensors. However, the effectiveness of this technology in reducing false alarms in clinical settings remains unclear. This study assesses and compares the motion resistance of Analog SpO2 Technology of two devices in the market.
Methods
Thirty healthy adults underwent controlled experiments (Control, Linear Motion, Angular Motion) using two pulse oximeters. Linear Motion tested hand displacement impact, while Angular Motion involved rhythmic hand motions at 120 bpm and 160 bpm.
Results
Both devices performed similarly in Control, with no disruptions. In Linear Motion, mild disruptions occurred, but no significant differences in SpO2 readings or alarms. Angular Motion at 120 bpm showed stability with no alarms. At 160 bpm, Device B (Biolight Analog SpO2) had fewer technical alarms but more SpO2 alarms than Device A (Masimo Analog SpO2).
Conclusions
Analog SpO2 exhibited motion resistance under static, linear and continuous waving angular motion up to 120 bpm and 160 bpm, but alarms occurred at 160 bpm with continuous tapping angular motion. These findings signify non-inferiority of either device in clinical settings. Further studies should include patients with cardiovascular and/or respiratory diseases.
Trial Registration
The study was submitted to and approved by the Biolight Ethics Committee (S0723), and written informed consent from all participants was obtained.
Graphical abstract

Introduction
A pulse oximeter is a non-invasive tool used to measure arterial blood oxygen saturation [1]. It works by shining light through the skin with its photodiodes and detecting it with a sensor that records light absorption as a function of time and can measure the percentage of hemoglobin bound to oxygen [2]. This lets healthcare providers assess the patient’s oxygenation status, making pulse oximetry one of the most important devices in the medical industry [3]. For patients of all ages, healthcare providers rely on these devices to diagnose hypoxemic events and observe treatment responses, which helps them make informed clinical decisions [4,5,6,7,8,9,10,11]. The pulse oximeters are useful devices, but sometimes their values are not very accurate, which can be due to a number of factors, including motion [12]. Previous studies have shown that motion has caused SpO2 (SpO2 = peripheral oxygen saturation) readings to be lower than actual readings and caused unnecessary alarms [13,14,15]. These alarms can lead to alarm fatigue, hindering attention when required. Engineers and researchers have worked together to remove these motion artifacts, which claim them to be very effective [16, 17]. Analog SpO2 technology measures and transmits the data in analog waveforms, which allows the devices to filter motion artifacts with proprietary algorithms and have proven to be better than digital sensors [17]. Even though this technology has existed for a while, newer publications on alarm fatigue still mention lots of false alarms in clinical settings, including technical and non-actionable alarms [18]. These alarms can be caused by many monitored parameters. Whether these alarms are caused by Analog SpO2 Technology or lack of adaptation of this technology is unknown. We hypothesize that Analog SpO2 technology could be more resistant to motion and thus decrease false alarms. The objective of this study was to assess and compare the extent of motion resistance of Analog SpO2 Technology among devices that currently offer this technology.
Method
Patient characteristics
This was a single-center, experimental study performed at the laboratory of Biolight from March 30, 2023 to April 21, 2023. In this study, we recruited a total of 30 adult subjects (15 males and 15 females) with ages ranging from 24 to 52 years. All participants were adults and in good health with various skin-tones, with no known cardiovascular or respiratory conditions, as that could significantly affect their SpO2 readings, and would have made it difficult to distinguish the cause of alarms, as this study’s objective was to assess motion resistance. Additionally, participants were required to abstain from consuming caffeine or alcohol for at least 12 h prior to the experiments to minimize potential confounding factors. Measurements were taken under these controlled environmental conditions to minimize the impact of extraneous factors on SpO2 readings.
Ethics approval and consent to participate
The study was submitted to and approved by the Biolight Ethics Committee (S0723), and written informed consent from all participants was obtained.
Sample size calculation
The minimal sample size was calculated according to our unpublished preliminary study, which showed that movement caused a 34% chance of SpO2 alarms with digital SpO2 devices used in the hospital compared to 8% alarms with Device A and Device B. We set α = 0.05, 1-β = 0.9, and used the “Sample Size Calculator” to calculate the sample size online [19]; the total sample size was 24. Considering the loss to follow-up rate of 25%, it was estimated that 30 cases need to be included.
Equipment used
-
1.
Biolight AnyView P22 Patient Monitor (Patient Monitor, Guangdong Biolight Meditech Co., Ltd., Zhuhai, People’s Republic of China).
-
2.
Biolight SpO2 module (Module for Patient Monitor, Guangdong Biolight Meditech Co., Ltd., Zhuhai, People’s Republic of China).
-
3.
Masimo SpO2 module (Module for Patient Monitor, Masimo Corporation, Irvine, USA).
-
4.
Biolight adult reusable analog SpO2 sensor 15-199-0359 (Accessory for Patient Monitor, Guangdong Biolight Meditech Co., Ltd., Zhuhai, People’s Republic of China).
-
5.
Masimo SpO2 adult reusable Finger Clip Sensor RD SET DCI REF 4050 (Accessory for Patient Monitor, Masimo Corporation, Irvine, USA).
-
6.
Masimo RD Rainbow SET MD20-12 Patient cable REF 4073(Accessory for Patient Monitor, Masimo Corporation, Irvine, USA).
-
7.
120 cm Camera railing with slider and camera or phone mounting attachments (Camera Railing and mount, Sutesheyingqicai Company, Shenzhen, China).
-
8.
Apple iPhone XS Max (with Metronome App and Camera App) (Apple, Cupertino, USA).
-
9.
Zip ties (Shaodeng Plastic Company, Leqing, China).
-
10.
Paper Scissors (Deli, Ninghai, China).
-
11.
Surgical tape (Haishi, Qingdao, China).
-
12.
Soft Velcro cable management tape (Xinhui, Shanghai, China).
-
13.
1-meter-long centimeter-scale ruler (Xiaobangshou, Suzhou, China).
-
14.
Large geometry set protractor (Jinsihuo, Taizhou, China).
Outcomes
Technical and SpO2 alarms were considered the primary outcome, and SpO2 Values and Disruptions see Fig. 1 in Plethysmography Waveforms were considered secondary outcomes. Disruptions in Plethysmography Waveforms were defined as none, mild, moderate, severe, and flatline. The acceptable thresholds of non-inferiority were determined by not sounding significantly more alarms, and not having a significantly large difference in SpO2 values or waveform disruptions. This significance is equivalent to statistical significance.
Statistical analysis
The differences in the primary and the secondary outcomes were assessed for significance by comparing means between the results of the two devices, one-sided p values were considered. The differences were considered significant if p < 0.05. The effect size (Cohen’s d) for the comparison of the two devices was calculated to quantify the magnitude of the difference, with a value of 0.2 indicating a small effect. To assess agreement between the two devices; Bland-Altman plot was generated using the SpO2 percentage values, from the Angular Motion experiment at 120 bpm and 160 bpm with both waving and 2-tap motion data separately, of the two devices. Statistical analyses were performed with SPSS 26.0 (IBM, Chicago, IL, USA).
Procedure
We performed three experiments: a Control Experiment, a Linear Motion Experiment, and an Angular Motion Experiment.
Control experiment
In the control group, SpO2 and the appearance of Plethysmography waveforms were monitored for 10 min on both Device A (Masimo Analog SpO2) and Device B (Biolight Analog SpO2) while keeping the hand still. First, Device A’s Probe was placed on the middle finger, and Device B’s Probe was placed on the index finger. After 5 min, the probes were switched; Device B’s Probe was placed on the middle finger, and Device A’s Probe was placed on the index finger.
Linear motion experiment
SpO2 readings during the Linear Motion Experiment included static readings and dynamic readings. Static readings were taken right before the hand was displaced when readings stabilized. The dynamic readings were the lowest readings within 15 s after the motion.
Two types of alarms were recorded: SpO2 alarms, the lower limit set was 90%, and technical alarms. If the alarm sounded within 15 s after the motion, it was recorded as a yes; otherwise, it was recorded as a no. This was later translated into statistical values: a yes translated into a 1, and a no translated into a 0.
In the Liner Motion Experiment, there were three groups. The groups were named “25 cm”, “50 cm”, and “75 cm”. In each of these groups, participants moved their hands with Biolight’s and Masimo’s probes attached for 25 cm, 50 cm, and 75 cm, respectively. To ensure that they all moved their hands at the same velocity, we used a camera railing with a heavy slider to calibrate the velocity and be used as a visual reference. The slider moved under its weight when the camera railing was placed at an inclined angle. This slider’s motion was taken as reference and the participants moved their hands along with the slider without touching it to match its speed.
The displacement was adjusted by measuring two points on the railing with a ruler and marking them with zip ties. The velocity was increased by increasing the height of one end of the railing. This height was determined by the angle of the inclined railing. This angle was set using a protractor. A video of the camera mount sliding down the railing was recorded using the phone’s camera. This was done to calculate the velocity by counting frames that showed motion of the camera mount between the two marked points. The velocity was confirmed by repeating the motion and recording it three times, and the most reliable readings of velocity were recorded.
After the motion, to avoid any unrelated alarms, the hand was kept steady for another 15 s to observe for any changes in the waveform or alarms sounded, and the lowest SpO2 readings within 15 s of the motion were recorded. Each length and velocity subset were repeated 10 times and readings were recorded. Device A and Device B probes were placed on the middle and index fingers, respectively, for the first five readings, and for the next five readings, the placement of the probes was switched. See Fig. 2.
Linear Motion Experiment Setup. The left side shows a camera railing with a slider attached and the researcher’s hand hovering above it with a Biolight SpO2 sensor on the index finger and a Masimo SpO2 sensor on the middle finger. The patient monitor on the right is a Biolight Anyview P22 patient monitor, displaying Masimo SpO2 above (label on screen: SpO2L) and Biolight SpO2 below (label on screen: SpO2)
Angular motion experiments
In this experiment, the participants moved their arms in an angular motion. To keep the motion consistent among all the participants, their arms were tied to a large compass with Velcro ribbons attached to it to secure the arm in place and allow for a consistent angular motion of 90° at the pivot point. See Fig. 3.
Angular Motion Experiment Setup. A large compass is tapped to a flat table surface with a protractor placed right on its pivot point, marking a 90° angle. There are two Velcro ribbons firmly attached to the compass’s left shaft to secure the arm to it. A paper sheet was placed on the surface to reduce friction, and water bottles were placed to mark the boundaries
The participants followed a metronome app that sounded a continuous rhythm of 120 bpm and 160 bpm. Each pace was considered a subgroup, 120 bpm and 160 bpm subgroups. We choose rhythms of 120 bpm and 160 bpm for this experiment because in our preliminary experiment we had 4 subgroups of rhythms 40 bpm, 80 bpm, 120 bpm and 160 bpm; there was no statistical difference in the 40 bpm, 80 bpm and 120 bpm subgroups. Therefore, we chose the fastest rhythm among these to be tested against 160 bpm. The data was recorded by the monitor, and timestamps were recorded on paper to later access the data from the monitor. The trend data for the Angular Motion Experiments was exported directly from the monitor with the help of the software engineer and was then filtered and classified by timestamps. The filtration involved removing empty data cells when the monitor was not making readings during rest or between participants. Filtration also meant only keeping the 60 s of data during the experiment; any values after the experiment were removed. The data was in intervals of 5 s. So, there were 12 readings for each subgroup.
Waving motion group
In the Waving Motion group, the participants waved their hands, rotating at the elbows by about 90 degrees in a rhythmic motion for 60 s at 120 bpm and then for another 60 s at 160 bpm. However, the actual movement of the hand was slightly less than 90 degrees since 90 degrees was a bit uncomfortable for participants to do. Two areas were marked on the table where the participants could float their hands over while keeping their elbows on the table. The participants were asked to rest for 3 min before performing each subgroup experiment.
2-tap motion group
In the 2-Tap Motion group, the process was the same as the Waving Motion group, except that with each beat, the participants had to softly tap on the surface of the table along with the waving motion.
Confounding factors
To control for potential external influences, all measurements were conducted under standardized conditions. Subjects were instructed to abstain from caffeine, alcohol, and medications that could affect cardiovascular or respiratory function before the experiment. Each subject was given time to rest upon arrival to minimize physiological variation due to nervousness or recent activity. To prevent light interference, room lights were turned off, and temperature was main tained at a consistent level for all trials. The fingers on which the probes were connected were also switched in a way that each probe was on each of the two fingers for 50% of the time. Participants were intentionally selected to include varied skin tones, reflecting the device’s applicability across diverse real-world populations.
Results
We recruited a total of 30 participants; however, 2 participants were unable to comprehend the instructions clearly and randomly stopped in the middle of the experiment or started talking. Therefore, their results were omitted, and 4 participants who participated later changed their minds and revoked their consent to let us publish their experimental results.
The results were consistent across all participants, suggesting that the control measures effectively minimized variability. Furthermore, the inclusion of participants with varied skin tones supports the device’s robustness and reliability across a diverse population.
According to the results of The Control Experiment, there were no waveform disruptions, no technical or SpO2 alarms, and the SpO2 values of both Device A and Device B had no significant differences regardless of the finger the sensor was attached to. See Table 1.
According to the linear motion experiment data, as the displacement increased from 0.25 m to 0.75 m, the waveform disruption increased from 0.26 to 1.02 for Device A and 0.36 to 0.68 for Device B. The mean waveform disruption of Device A and Device B was 0.65 and 0.53, respectively. Both indicate None to Mild waveform disruptions. Static and Dynamic SpO2 values of both Device A and Device B were similar, with no significant differences. No technical or SpO2 alarms were sounded by either Device A or Device B throughout this experiment. See Table 2.
With respect to velocity, the waveform disruptions peaked at 1.5 for Device A, which is between the Mild to Moderate range, and at 0.9 for Device B, which is between the None to Mild range. The increase in disruptions did not show a clear association with velocity. For Device A, the waveform disruptions increased with a velocity of up to 0.43 m/s; however, it went back down to 0.2 at 0.45 m/s and all the way to 0.00 at 0.47 m/s and then went back up to 1.00 at 0.66 m/s. For Device B, the waveform disruptions were higher at both lower and higher velocities of 0.31 m/s, 0.32 m/s, 0.33 m/s, 0.52 m/s, 0.66 m/s, and 0.77 m/s while lower at other velocities. See Table 2.
According to the results of the Angular Motion Experiment, in the 120 bpm subgroup, Device B’s SpO2 values remained stable between 95 and 98 throughout the experiment. In contrast, Device A’s SpO2 values dropped from 98.5 to 94% after 8 s but recovered to 98% after another 10 s. In the 160 bpm subgroup, Device B’s SpO2 values gradually decreased below 97–94.4%, whereas Device A’s SpO2 values remained stable at 98%. Device A showed slightly higher levels of waveform disruptions in both subgroups when compared to Device B. However, the difference was not significant. The waveform disruption was moderate in both. See Fig. 4.
Waving motion group
Regarding the Technical and SpO2 Alarms, the results for the Waving Motion Group showed no alarms for both devices for the 60-second duration in both the 120 bpm and 160 bpm subgroups.
2-tap motion group
In the 2-Tap Motion Group, there were no alarms for the entire 60-second duration in the 120 bpm subgroup. However, in the 160 bpm subgroup of the 2-Tap Motion Group, in the first 20 s, Device A sounded 2 technical alarms compared to Device B, which sounded 1, whereas Device B sounded 4 SpO2 Alarms compared to Device A, which sounded 1. From 20 to 60 s, Device A sounded 6 technical alarms, whereas Device B sounded 17 SpO2 Alarms during this time. During this duration of 20–60 s, Device A did not sound any SpO2 alarms, whereas Device B did not sound any Technical Alarms. See Table 3.
Discussion
The primary outcome of technical and SpO2 alarms showed no significant differences up in all experiment groups and subgroups except in the 160 bpm 2-tap motion experiment. The SpO2 values and waveform disruptions showed no significant differences. We conducted a total of three experiments with the objective of identifying the anti-motion resistance capabilities of Analog SpO2 Devices. We tested two devices for motion resistance and compared them to each other. The purpose of these experiments was to evaluate the clinical use of Analog SpO2 Devices under motion.
The control experiment
According to the results of the first experiment, The Control Experiment, when there was no motion, there were no waveform disruptions, and no technical or SpO2 Alarms. This experiment proved that the setup was stable, and both devices performed nominally.
Linear motion experiment
The Linear Motion Experiment was designed to test the effect of linear displacement over a fixed distance that a patient would move their hand in the hospital. The purpose of having a fixed defined distance and an accelerating slider on rails was to test the device up to its limits; however, with our testing, we were unable to reach the limits with this controlled movement. Initially, during our preliminary testing phase, we had planned to use a simulator to minimize the number of variables as we could fix the simulator to the camera slider. However, we realized that the setup wouldn’t work as the simulator did not produce any wave disruptions or changes in readings; it could not simulate a real human hand. Therefore, we decided to do the experiment with real human participants who volunteered to participate. With all the liner motion experimentation, we found that the Analog SpO2 readings of both devices were comparable, there were no alarms, and the waveforms of Device B were more stable than those of Device A; however, there was no significant difference. These experiments were quite challenging for both devices, yet both devices displayed anti-motion resistance capabilities.
Angular motion experiments
From The Linear Motion experiments, we believe that these experiments, although they do represent a certain kind of hand motion, the actual motion of the hand might also involve some angular motion since the hand is attached to the forearm and pivots at the elbow, creating an angular motion. These experiments also proved that even with a rigorous continuous motion of up to 160 bpm, neither device sounded any alarms, and the readings were almost identical as long as the hand didn’t bump into anything, which is what we saw in The Waving Motion group. Even when it bumped into things, the readings were stable, and no alarms sounded at 120 bpm with both devices.
In these experiments, we collected the data in three phases: 0–20 s, 20–40 s, and 40–60 s. If a device started sounding the alarm in the first phase and ended in the last phase, that counted as 3 alarms instead of 1. In the 160 bpm experiment, Device A sounded more technical alarms, whereas Device B sounded more SpO2 alarms. In total, Device A sounded 8 technical alarms, while Device B sounded 1. Device A sounded 1 SpO2 alarm, whereas Device B sounded 21. It is to be noted that when Device A sounded technical alarms, it was not reading SpO2 values accurately. Therefore, the technical alarms, and thus it means that it could not sound SpO2 alarms during that period. Device B’s alarms could either be misreadings or an actual decrease in SpO2 due to decreased perfusion; it is hard to judge the exact reason for the alarms.
Most of Device B’s SpO2 alarms started sounding in the 20–40 s phase and ended either after the experiment ended or in the 40–60 s phase. This means that for the first 20 s, there were fewer alarms from both devices, which is what we believe exceeds the maximum duration of such continuous rigorous motion for a patient in the hospital. The 160 bpm 2-tap angular motion is very fast-paced, and we believe the reason for most of the alarms might not be the motion but the tapping on a hard surface. Tapping at such a high pace could misalign the sensor, causing it to sound alarms.
Clinical implications
The findings of this study demonstrate that Analog SpO2 Pulse Oximetry exhibits motion resistance up to limits significantly higher than those encountered in a typical hospital or clinical setting during routine use. This suggests that this technology can be considered motion-resistant for daily clinical use. However, it is imperative for healthcare providers to maintain a nuanced understanding of these devices’ limitations, as occasional motion artifacts can still occur. Waveform disruptions generally do not trigger alarms, it is when the device is unable to monitor the SpO2 value of the patient, that it triggers an alarm. In our results Device A considered this a technical issue and Device B considered it an SpO2 alarm. There are essentially two types of alarms: technical or physiological. Technical alarms sound when there is a technical issue with the device, and it is unable to monitor the physiological parameter for any reason, which could be a faulty sensor, when the device detects motion, when the device detects too much disturbance in the signal, or other technical reasons. These alarms are generally low priority alarms and make less noise compared to physiological alarms. Physiological alarms such as SpO2 alarms sound when the device is able to monitor the physiological parameters nominally, however, it detects that the physiological measurement is above or below a certain limit that is defined by the user and generally has a default value set by the manufacturer. These alarms are designed with in accordance with standards set by the International Organization for Standardization (ISO), specifically IEC 60601-1-8:2006 standard [20]. When a technical alarm sounds, it means the patient is not being monitored for now due to a technical reason, this does not mean that an abnormal physiological event cannot occur at this time. Therefore, the patient needs to be monitored manually until the issue that caused the technical alarm is resolved. When a physiological alarm sounds, that means the patient’s physiological parameter has exceeded the defined limit and the healthcare person on duty must attend to the patient immediately. Regardless of the type of the alarm, when the alarm sounds, the physician or nurse on call should respond in a timely manner. However, many times some of these alarms are considered false alarms since they might be caused as a result of a technical issue. This analog technology however could potentially reduce the number of false alarms significantly.
Minimizing alarm fatigue
Alarms triggered by motion artifacts can result in an overwhelming number of false alarms leading to many hospitals completely turning the alarms off or getting somewhat immune to the sound. Analog SpO2 technology could minimize the alarms caused by motion artifacts and allow healthcare providers to focus on alarms that matter. Improving Patient Safety and reliability of the device in the healthcare setting. Healthcare professionals should also read user manuals that come with the devices in order to make sure of the proper usage of the devices. Some things that can help reduce false alarms are making sure that the surface of the finger on which the sensor is attached to is not dirty, over-pigmented, tattooed, injured, nail polished, cold, elevated, wet, on the same side as the blood pressure cuff is on, too large or too small for the probe, excessively in motion, exposed to bright lights other than from the probe, tucked under the body, or have reduced blood perfusion. The exact instructions on the proper usage of the device can differ and therefore instructions for a specific device are often mentioned in the user manuals provided by manufacturers, which healthcare professionals should read carefully before use.
Future research
The study highlights the need for further research into technologies and techniques that can further reduce the number of false alarms triggered by non-medical events. Future studies should also test these technologies on cardiac disease and/or respiratory disease patients.
Limitations
Most alarms occurred when the subject’s hands were cold. We did not check the temperature of the hands before the experiment; after the experiment, we touched the hands of the participants and realized they were cold. However, we could not heat the hands and re-record the readings as this was not part of the design of the experiment. However, we do believe that perfusion could have been improved by heating the hands a little, and the results of the experiment might have had fewer alarms. Another limitation is the small sample size and single-center study; the results might have been more generalizable with more participants from different centers. However, the participants in this study did have different ethnic backgrounds with varying complexions of skin, as this has been shown to affect the results of pulse oximetry [21, 22]. Another weakness of this study is that it was only performed on adults, because of how hard it would have been for children to comply with since the instructions were quite complex. Lastly, this study only compared two technologies using one monitor. We contacted some other brands, but they didn’t seem very keen to lend us their equipment for testing.
Conclusions
In conclusion, our study found that Analog SpO2 was motion resistant under static, controlled linear motion conditions, and continuous rapid angular motion of up to the pace of 120 bpm. However, alarms were noted with a pace of 160 bpm. These findings emphasize the importance of healthcare provider awareness regarding device limitations and the need for further research to enhance monitoring devices for improved clinical utility.
Data availability
The datasets generated and/or analyzed during the current study are not publicly available due to data format being unable to upload to repositories but are available from the corresponding author on reasonable request.
Abbreviations
- bpm:
-
Beats per minute; musical beats for rhythm generated by a metronome phone application
- Device A:
-
Masimo Analog Pulse Oximeter
- Device B:
-
Biolight Analog Pulse Oximeter
- SpO2 :
-
Pulse Oximetry obtained via Blood Oxygen Saturation
References
Nasr VG, DiNardo JA. Pulse oximetry. Pediatr Rev. 2019;40(11):605–8.
Leppänen T, Kainulainen S, Korkalainen H, Sillanmäki S, Kulkas A, Töyräs J, Nikkonen S. Pulse oximetry: the Working Principle, Signal formation, and applications. Adv Exp Med Biol. 2022;1384:205–18.
Tekin K, Karadogan M, Gunaydin S, Kismet K. Everything about pulse oximetry-part 1: history, principles, advantages, limitations, inaccuracies, cost analysis, the Level of Knowledge about Pulse Oximeter among clinicians, and pulse Oximetry Versus tissue oximetry. J Intensive Care Med. 2023;38(9):775–84.
Wick KD, Matthay MA, Ware LB. Pulse oximetry for the diagnosis and management of acute respiratory distress syndrome. Lancet Respir Med. 2022;10(11):1086–98.
Martin GR, Ewer AK, Gaviglio A, Hom LA, Saarinen A, Sontag M, Burns KM, Kemper AR, Oster ME. Updated strategies for pulse oximetry screening for critical congenital heart disease. Pediatrics 2020, 146(1).
Shindova MP, Belcheva-Krivorova A, Taralov Z. Pulse oximetry in paediatric dentistry. Folia Med (Plovdiv). 2022;64(2):202–6.
Abbas A, Ewer AK. New born pulse oximetry screening: a global perspective. Early Hum Dev. 2021;162:105457.
Rackley CR. Monitoring during mechanical ventilation. Respir Care. 2020;65(6):832–46.
Shepard LN, Berg RA, Morgan RW. Pulse oximetry plethysmography: a new approach for physiology-directed CPR? Resuscitation. 2021;169:198–200.
Ekström M, Engblom A, Ilic A, Holthius N, Nordström P, Vaara I. Calculated arterial blood gas values from a venous sample and pulse oximetry: clinical validation. PLoS ONE. 2019;14(4):e0215413.
Prosperi P, Verratti V, Bondi D, Spacone A. On pulse oximetry and hypoxia. Respir Physiol Neurobiol. 2023;315:104111.
Bickler P, Tremper KK. The pulse oximeter is amazing, but not perfect. Anesthesiology. 2022;136(5):670–1.
Dormishian A, Schott A, Aguilar AC, Bancalari E, Claure N. Pulse oximetry reliability for detection of hypoxemia under motion in extremely premature infants. Pediatr Res. 2023;93(1):118–24.
Mark N, Lyubin A, Gerasi R, Ofir D, Tsur AM, Chen J, Bader T. Comparison of the effects of Motion and Environment conditions on Accuracy of Handheld and Finger-based pulse oximeters. Mil Med. 2021;186(Suppl 1):465–72.
Giuliano KK, Bilkovski RN, Beard J, Lamminmäki S. Comparative analysis of signal accuracy of three SpO(2) monitors during motion and low perfusion conditions. J Clin Monit Comput 2023.
Majeed IA, Jos S, Arora R, Choi K, Bae S. Motion artifact removal of Photoplethysmogram (PPG) Signal. Annu Int Conf IEEE Eng Med Biol Soc. 2019;2019:5576–80.
Barker SJ. Motion-resistant pulse oximetry: a comparison of new and old models. Anesth Analg. 2002;95(4):967–72. table of contents.
Ruskin KJ, Hueske-Kraus D. Alarm fatigue: impacts on patient safety. Curr Opin Anaesthesiol. 2015;28(6):685–90.
Sample Size Calculator. [https://clincalc.com/Stats/SampleSize.aspx]
Medical electrical equipment. — General requirements for basic safety and essential performance — Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems. In: IEC 60601-1-8:2006(en). Edited by Standardization IOf.
Jamali H, Castillo LT, Morgan CC, Coult J, Muhammad JL, Osobamiro OO, Parsons EC, Adamson R. Racial disparity in Oxygen Saturation measurements by pulse oximetry: evidence and implications. Ann Am Thorac Soc. 2022;19(12):1951–64.
Shi C, Goodall M, Dumville J, Hill J, Norman G, Hamer O, Clegg A, Watkins CL, Georgiou G, Hodkinson A, et al. The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis. BMC Med. 2022;20(1):267.
Acknowledgements
Thanks to Guangdong Biolight Meditech Co., Ltd. for lending us the Masimo module and the P1 transport monitor with the P22 Patient Monitor.
Funding
This study was not funded.
Author information
Authors and Affiliations
Contributions
T.Y. analyzed the data, wrote the manuscript, and supervised the study. Y.Liu. helped with data analytics. F.H.C. collected the data. Y.Li. extracted the data from the patient monitor and helped understand it. M.S.M. designed the study, collected the data, and analyzed the data.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
The study was submitted to and approved by the Biolight Ethics Committee (S0723), and written informed consent from all participants was obtained.
Consent for publication
Written informed consent for publication was obtained from all participants.
Competing interests
One of the analog technologies mentioned in this paper was developed by Guangdong Biolight Meditech Co., Ltd. Yong Liu, CaiHua Feng, Yong Li, and Muhammad Saqib Mudabbar are associated with this organization.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Yang, T., Liu, Y., Cai, F. et al. Motion resistance in peripheral oxygen saturation monitoring using Biolight Analog SpO2 compared to Masimo SpO2: a non-inferiority study. BMC Anesthesiol 24, 430 (2024). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12871-024-02823-z
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12871-024-02823-z