Skip to main content

Table 2 Linear Motion Experiment. Data is shown as mean ± standard deviation (n = 24)

From: Motion resistance in peripheral oxygen saturation monitoring using Biolight Analog SpO2 compared to Masimo SpO2: a non-inferiority study

Linear Motion in terms of Displacement Value

Displacement (m)

Device A1 Waveform Disruption (0–4)

Device B2 Waveform Disruption (0–4)

Device A SpO2 (%±S)

Device B SpO2 (%±S)

Device A Technical Alarms (n ± S)

Device B Technical Alarms (n ± S)

Device A SpO2 Alarms (n ± S)

Device B SpO2 Alarms (n ± S)

0.25 m (x̄±s)

0.3 ± 0.4

0.4 ± 0.6

99.3 ± 0.7

99.6 ± 0.5

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.5 m (x̄±s)

0.7 ± 0.7

0.5 ± 0.6

99.4 ± 0.6

99.9 ± 0.3

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.75 m (x̄±s)

1 ± 0.9

0.7 ± 0.7

99.5 ± 0.5

99.8 ± 0.4

0 ± 0

0 ± 0

0 ± 0

0 ± 0

Linear Motion in terms of Finger used

Finger

Device A Waveform Disruption (0–4)

Device B Waveform Disruption (0–4)

Device A SpO2 (%±S)

Device B SpO2 (%±S)

Device A Technical Alarms (n ± S)

Device B Technical Alarms (n ± S)

Device A SpO2 Alarms (n ± S)

Device B SpO2 Alarms (n ± S)

Index (x̄±s)

0.5 ± 0.7

0.5 ± 0.7

99.4 ± 0.6

99.8 ± 0.4

0 ± 0

0 ± 0

0 ± 0

0 ± 0

Middle (x̄±s)

0.8 ± 0.8

0.5 ± 0.6

99.4 ± 0.6

99.7 ± 0.5

0 ± 0

0 ± 0

0 ± 0

0 ± 0

Linear Motion in terms of Velocity

Velocity (m/s)

Device A Waveform Disruption (0–4)

Device B Waveform Disruption (0–4)

Device A SpO2 (%±S)

Device B SpO2 (%±S)

Device A Technical Alarms (n ± S)

Device B Technical Alarms (n ± S)

Device A SpO2 Alarms (n ± S)

Device B SpO2 Alarms (n ± S)

0.26

0.2 ± 0.4

0.1 ± 0.3

98.9 ± 0.3

99.6 ± 0.7

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.31

0.6 ± 0.5

0.9 ± 0.7

100 ± 0

100 ± 0

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.32

0.7 ± 0.5

0.9 ± 0.7

99.3 ± 0.5

99.9 ± 0.3

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.33

1 ± 0.8

0.7 ± 0.7

99.6 ± 0.5

99.8 ± 0.4

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.37

0.3 ± 0.5

0.1 ± 0.3

98.9 ± 0.6

99.3 ± 0.5

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.4

1.3 ± 0.8

0.6 ± 0.7

99.4 ± 0.5

99.8 ± 0.4

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.43

1.5 ± 1

0.2 ± 0.4

99 ± 0

99.8 ± 0.4

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.45

0.2 ± 0.4

0.6 ± 0.7

98.5 ± 0.5

99.3 ± 0.5

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.47

0 ± 0

0.1 ± 0.3

100 ± 0

99.9 ± 0.3

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.5

0.6 ± 0.7

0.5 ± 0.5

99.1 ± 0.6

100 ± 0

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.52

0.8 ± 0.4

0.8 ± 0.8

99.6 ± 0.5

100 ± 0

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.58

0.2 ± 0.4

0.3 ± 0.5

99.4 ± 0.7

99.8 ± 0.4

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.65

0.6 ± 0.8

0.4 ± 0.7

99.9 ± 0.3

99.9 ± 0.3

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.66

1 ± 1.3

0.8 ± 0.9

99.6 ± 0.5

99.6 ± 0.5

0 ± 0

0 ± 0

0 ± 0

0 ± 0

0.71

0.8 ± 0.6

0.9 ± 0.6

99.9 ± 0.3

99.9 ± 0.3

0 ± 0

0 ± 0

0 ± 0

0 ± 0

  1. 1Device A = Masimo Analog SpO2, 2Device B = Biolight Analog SpO2