- Research
- Open access
- Published:
Reliability of nociceptive monitors vs. standard practice during general anesthesia: a prospective observational study
BMC Anesthesiology volume 25, Article number: 51 (2025)
Abstract
Background
Inadequate or excessive nociceptive control during general anesthesia can result in significant adverse outcomes. Using traditional clinical variables, such as heart rate, systolic blood pressure, and respiratory rate, to assess and manage nociceptive responses is often insufficient and could lead to overtreatment with both anesthetics and opioids. This study evaluated the feasibility and effectiveness of three nociception monitoring techniques Nociception Level Index (NOL), Skin Conductance Algesimeter (SCA) and heart rate monitoring in patients undergoing image-guided, minimally invasive abdominal interventions under general anesthesia.
Method
This prospective observational study collected data from 2022 to 2024. All patients were anesthetized according to the departmentâs routine, and predetermined events were recorded. Two commercially available nociception monitors, the PMD-200 from Medasense (NOL) and PainSensor from MedStorm (SCA), were used, and their data were collected along with various hemodynamic parameters. The three nociception monitoring techniques were compared during predetermined events.
Result
A total of 49 patients were included in this study. NOL and SCA demonstrated higher responsiveness than HR for all events except for skin incision. The comparison of the values above and below the threshold for each nociceptive stimulus showed significance for all measurements using the SCA and NOL. However, using HR as a surrogate for nociception with a threshold of a 10% increase from baseline, the difference was significant only at skin incision. There was no variation in the peak values attributable to differences in patientsâ age. Weight was a significant predictor of the peak NOL values.
Conclusion
NOL and SCA demonstrated superior sensitivity and responsiveness to nociceptive stimuli compared to HR, effectively detecting significant changes in nociceptive thresholds across various stimuli, although responses during skin incision showed no such advantage.
Trial registration
Clinical trial - NCT05218551.
Background
Evidence links general anesthesia to post-operative cognitive dysfunction (POCD) [1, 2], with nociception and anesthetic depth being key factors. Analgesics, especially opioids, mitigate nociception but their intra- and post-operative effects are not fully understood. Reliable nociceptive monitoring during surgery remains elusive, highlighting the need for better tools to tailor analgesia [3].
Inadequate or excessive nociceptive control can lead to adverse outcomes, such as Remifentanil-induced hyperalgesia and prolonged post-operative pain [4,5,6]. Individual variability in opioid response, influenced by genetics, underscores the need for personalized approaches [7,8,9]. With Europeâs elderly population expected to double by 2050 [1], refined anesthesia practices are critical to ensure safety and efficacy for this vulnerable group.
Heart rate is commonly used to assess nociception under anesthesia, but it is influenced by non-nociceptive factors, making subjective interpretation necessary [10,11,12]. Over-reliance on clinical parameters like heart rate, blood pressure, and respiratory rate may lead to overtreatment with anesthetics or opioids, affecting hemodynamic stability and complicating management [12,13,14].
Recent nociception monitors, such as the Medasense PMD-200 and MedStorm PainSensor utilizes parameters such as heart rate and/or skin conductance to objectively quantify nociceptive responses [15]. The PMD-200 provides a Nociception Level Index (NOL) ranging from 0 to 100, with 25 as the treatment threshold [16]. In contrast, the PainSensor, exemplified by the Skin Conductance Algesimeter (SCA) index, employs either a treatment threshold of 0.06 peaks/second or a scale ranging from 0 to 10, with a threshold of â„â3. These devices show potential but require further validation, particularly in complex patient populations [17, 18].
This study assessed the feasibility and effectiveness of three nociception-monitoring techniquesâNOL, SCA, and heart rate monitoringâin patients undergoing image-guided minimally invasive abdominal interventions under general anesthesia.
Aim
To evaluate the effectiveness of different somatic nociceptive monitoring techniques and visualize their responses to various nociceptive stimuli during minimally invasive procedures under general anesthesia.
Patients and anesthesia methods
This was an observational study that included 49 patients. The assumption for the number of participants included in this observational study is based on calculations derived from the articles by Rantanen et al. [19] and Martini et al. [20].
Patients were screened for eligibility via medical records and asked to participate during the preoperative assessment visit at the hospital. All patients received written and oral information. Written informed consent was obtained from all participants. The trial conduct adhered to the Declaration of Helsinki and Good Clinical Practice guidelines.
The inclusion criteria specified adults aged 18 and older who had undergone minimally invasive abdominal interventions under general anesthesia with a small incision either in the abdomen or through the femoral artery or vein. Patients were excluded if they had an American Society of Anesthesiologists (ASA) risk classificationâ>âIV, chronic pain, or known neurological disorders [see Additional file 1].
All patients were anesthetized following the departmentâs routine, e.g., induction, PropofolÂź 1â3Â mg/kg IV and RemifentanilÂź administered via a target-controlled infusion pump (Minto model) [21] with an effect site target (Ce) of 5.0 ng/ml IV, and RocuroniumÂź 0.6Â mg/kg IV were used. During the skin incision, the Ce of Remifentanil was set to 4.0 ng/ml. Anesthesia was maintained using an infusion of Sevoflurane and Remifentanil.
All patients received vasopressors via IV infusion with Norepinephrine 0.04Â mg/ml and a crystalloid solution according to the department routine. PhilipsÂź Intellivue X3 was used for measuring vital parameters, including pulse oximetry, invasive or non-invasive blood pressure, 3-lead ECG, NMT, and SedLine (4-channel processed EEG MasimoÂź technology). Nociception was monitored before anesthesia induction. The MedStormÂź PainSensor was positioned on the patientâs left hand, and the MedasenseÂź PMD-200 (NOL) sensor was placed on the right or left hand.
Data collection and procedure
All the data were collected at a university hospital in Sweden between 2022 and 2024. All parameters, drugs, and predetermined events (jaw-thrust, laryngoscopy, endotracheal tube insertion, and skin incision) were recorded. The minimal alveolar concentration of sevoflurane (MAC), end-tidal carbon dioxide concentration (EtCO2), and non-invasive blood pressure measurements (SBP, DBP, and MAP) were recorded every 5Â min. The NOL Index was recorded every 5Â s, and the depth of anesthesia was recorded every 2Â s as spectral edge frequency (SEF-L and SEF-R), indicating that 95% of the EEG frequency was below the given Hz [22]. The SCA peaks/sec, pulse oximetry (SpO2), heart rate (HR), and invasive blood pressure (SBP, DBP, and MAP) were recorded every second.
The thresholds for commercial nociception monitors were based on manufacturer-specified values as the limits for initiating nociception treatment. The 10% increase in heart rate that we used for comparison with traditional clinical heart rate assessment via the monitor is not an established research-based limit but rather derived from clinical experience. However, a pilot study investigating similar stimuli showed that a 90% increase in NOL corresponded to a 12% increase in heart rate [23].
Hemodynamic and respiratory data were recorded using the Moberg CNS-200. Four devices (Medasense PMD-200, MedStorm PainSensor, Masimo SedLine, and Moberg CNS-200) were employed for data recording, and time synchronization among the devices was performed manually before each patient. All the recorded data were merged using a Python script in a Jupyter Notebook. The baseline to peak data (Table 2) were defined as the peak values recorded by the different monitors within ±â15 s of the event, occurring within aâ±â30-second window from the registered event time, to minimize potential errors related to manual event marking and time synchronization between devices.
Data collection began with patient preoxygenation and ended when the patient was transferred after extubation. Nociceptive monitors were blinded to all staff throughout the data collection process.
Calculation and statistics
The statistical methods utilized in this study included repeated measures ANOVA to assess changes over time, independent-samples t-tests to compare group means, and Pearsonâs correlation coefficient to evaluate the relationship between continuous variables. A Receiver Operating Characteristic (ROC) curve analysis was performed to evaluate the different monitoring techniques and to calculate the Area Under the Curve (AUC). All statistical calculations, analyses, and visualizations were performed using IBM SPSS Statistics for MacOS (Version 29.0.2.0, IBM Corp., Armonk, NY, USA) and RStudio (Version 2024.04.2â+â764, RStudio, PBC, Boston, MA, USA). Descriptive statistics and basic calculations were performed using Jupyter Notebook 6.4.8 (Python 3.0) and Microsoft Excel (version 16.85).
Result
All patients received the same level of analgesia based on department routines and variables such as weight, age, sex and height (Table 1).
The difference between the highest and baseline values was significant for all stimuli with all three measurement methods, except for SCA and HR during skin incision. However, NOL and SCA showed higher responsiveness than HR to all stimuli except skin incision (Table 2).
Comparison of the values above and below the threshold for each nociceptive stimulus showed significance for all measurements using the SCA and NOL. Hence, using HR as a surrogate for nociception with a threshold of a 10% increase from baseline, the difference was significant only at skin incision (Table 3).
Between the two nociception monitoring techniques, the SCA recorded more patients above the threshold during jaw-thrust and fewer during endotracheal tube insertion (Table 3). For skin incisions, both nociception monitors showed a significantly lower number of patients above the threshold. However, NOL showed 57% more patients above the threshold than SCA. During jaw-thrust and endotracheal tube insertion, the SCA showed the same number of patients above the threshold for both stimuli, while the HR and NOL varied. Regardless of the stimuli, heart rate detected a consistent number of patients over the threshold (Table 3).
A Pearson correlation was made for all stimuli combined and showed a significant relationship between peak values for NOL and HR (pâ=â<â0.001). A correlation analysis showed a strong and significant relationship for both NOL and SCA compared with HR (pâ=â<â0.001 versus pâ=â.002). When comparing the correlation between HR and NOL/SCA in peak values, the strongest predictor of elevated HR was an increase in SCA, although this was not significant (Bâ=â7.48, pâ=â.377). NOL peak values significantly predicted HR increase but had a lower beta value than SCA (Bâ=â0.16, pâ=â.025).
There were no significant differences in the peak values of NOL (pâ=â.216), SCA (pâ=â.313), or HR (pâ=â.458) based on sex. However, there was a significant difference in the increase from baseline to peak with NOL at the group level (pâ=â.015). There was no variation in peak values explained by age differences between patients in NOL (Bâ=âââ0.099, pâ=â.254), SCA (Bâ=â0.001, pâ=â.286), or HR (Bâ=âââ0.053, pâ=â<â0.001). Hence, weight was a significant predictor of peak values in the NOL technique used during nociceptive stimuli (Bâ=â0.21, pâ=â<â0.001).
The ROC curves for NOL, SCA and HR (all measured after stimulation) were calculated (nâ=â168) and are displayed in (Fig. 1). Sensitivity values at a specificity of 75%, along with the areas under the ROC curves, are provided in (Table 4). Among the variables, NOL demonstrated superior performance in distinguishing between noxious (jaw-thrust and intubation) and non-noxious events, achieving an AUC of 0.89 (95% CI, 0.86-0.96) (Table 4).
Showing receiver operating characteristics (ROC) curve of NOL, SCA and HR. ROC curve discrimination between nociceptive (jaw-thrust and intubation) and nonnociceptive stimuli at Remifentanil concentration (Ce 5/3 ng/ml). Sensitivity and specificity provided in Table 4
Discussion
This observational study compared three somatic methods of assessing nociception in response to known tissue stimuli. These stimuli were jaw-thrust, laryngoscopy, insertion of the endotracheal tube, and skin incision, which are known to trigger stress reactions and are perceived as nociceptive and have been used in previous research [20, 24, 25]. Our study used the same relative doses of analgesics and the same types of stimuli across individuals with varying medical histories. The three methods observed, which serve as surrogates for nociception, generally responded similarly yet exhibited some significant differences. Our results indicate and supports by other studies [12, 16, 26,27,28] that the heart rate is not superior and, in some cases, is even less effective than commercial nociception monitors.
Accordingly, all our patients were prescribed various medications to manage blood pressure and heart disease (Table 1). While research [29] suggests that adrenergic blockers may have a more pronounced effect on the nociceptive response to moderate knee pain compared to other antihypertensive drugs, this claim is contradicted by Zhou et al., [30] who found no evidence that beta blockers provide a clinically meaningful reduction in knee pain. In our case, the impact of these different medications on patients with cardiovascular conditions under general anesthesia, particularly concerning their nociceptive response, remains unclear.
This study also showed that monitoring nociception has advantages over the heart for assessing nociceptive stimuli. The nociceptive-extracted values were two to three times as many patients above the threshold compared to heart rate, indicating a stronger correlation with the stimulus in relation to heart rate only. This finding is supported by a study showing that SCA is superior to changes in HR and blood pressure in critically ill mechanically ventilated patients [31].
Both the NOL and SCA have been validated for their response to tissue stimuli and different levels of analgesia [16, 32]. Studies [17, 33] have shown that NOL-guided analgesia can reduce intraoperative opioid consumption and positively impact the experience of post-operative pain. However, a systematic meta-analysis [34] has also been conducted, which failed to demonstrate a definitive benefit of NOL in improving patient outcomes, highlighting the necessity for further research to validate its efficacy across varied clinical contexts. In this study, the SCA showed great similarity to the NOL, though significant differences were observed between the devices, with the SCA showing more consistent results and independence from patient demographics.
Studies have shown that individuals with obesity exhibit altered nociceptive processing, potentially due to ongoing and heightened inflammatory responses compared to individuals with normal weight [35,36,37,38]. Our study suggests that obesity, as a risk factor, may influence nociceptive mechanisms and serve as a predictor for elevated NOL values. This finding may be attributed to the NOL technique, which appears to demonstrate higher sensitivity and specificity compared to SCA and HR measurements. Furthermore, our results align with the findings of the review article by Martinez-Vazquez P et al. [12].
To some extent, all patients in this study showed an increase in somatic stress, for example, nociception levels, when exposed to stimuli, and half exceeded the threshold for treatment, as described by the respective manufacturer (NOL and SCA). All three nociception assessment methods used in this study identified these increases to varying degrees. During stimuli with higher doses of remifentanil and longer duration (jaw-thrust, laryngoscopy, and insertion of an endotracheal tube), HR was inferior to nociception monitors. For the skin incision, which was a shorter stimulus, all three methods behaved similarly. This could be because the tissue stimuli were not sufficiently intense to trigger a nociceptive response [39].
Interestingly, not all patients experienced adequate nociceptive relief despite the administration of high doses of remifentanil [6]. Approximately half of the patients exhibited nociceptive values indicative of the need for treatment, suggesting that these individuals were subjected to levels of surgical stress that could potentially have adverse effects on post-operative recovery and well-being [40]. This, in combination with the risks of overtreatment with opioids, shows the importance of closely monitoring these complex patients, as relying solely on HR does not seem to capture the variations in nociception level with the same veracity as NOL and SCA. With older age, more comorbidities, and polypharmacy, the influence on hemodynamic events grows, making it more difficult to correlate with the state of the nociceptive response. As individuals age, the prevalence of comorbidities and the complexity of polypharmacy increase, which heightens the impact on hemodynamic events [41]. This interplay complicates the ability to accurately correlate hemodynamic changes with nociceptive responses.
During anesthesia administered in our department, norepinephrine is used as the first-line vasopressor, consistent with recommendations highlighted in the literature [42]. However, the algorithm for the conductance technique was not influenced by calcium levels, as the method employed for assessing skin sympathetic nerve activity (SSNA) is inherently independent of calcium-mediated effects [43].
Anesthesia personnel tend to prioritize hemodynamic stability over analgesia when treating patients under general anesthesia. For instance, Carella M et al. [44] emphasized the importance of maintaining hemodynamic stability during anesthesia, noting that it often becomes the primary focus for anesthesiologists. The authors stated, âWe focused on the final systemic aspect of nociceptive stress response and considered hemodynamic stability to be our real challengeâ. While ensuring nociceptive balance is essential, immediate management of hemodynamic parameters often takes precedence, highlighting the complexity of using HR as a measure of nociception [45].
Additionally, research has highlighted the challenges in using HR as an indicator of nociception. A study by Meijer et al. [46] discussed the use of nociception monitors to guide opioid administration, aiming to optimize anesthesia and improve patient outcomes. The study found that while these monitors can aid in assessing nociception, relying solely on HR can be problematic due to its sensitivity to various factors beyond pain, such as physiological and pharmacological influences.
This study demonstrated the utility of nociceptive measurements, which may enable the identification of individuals with genetic variations related to increased tolerance or heightened sensitivity to analgesic medications with greater precision regarding their nociceptive responses [7]. Consequently, these individuals can be managed individually, ensuring the administration of an appropriate pain regimen with higher accuracy.
We have studied the sensitivity and effectiveness of various methods for describing nociceptive responses during general anesthesia. Our study includes a ROC analysis for different stimuli, incorporating various methods such as NOL, SCA and HR. The results suggest that composite methods perform well, and even better than HR alone, in detecting nociception. However, the aim of this study is to describe clinically available and effective method/s for nociception detection during general anesthesia that have already been validated for sensitivity and specificity.
Our findings, along with those of other studies [17, 47], underscore the challenges of achieving effective analgesia during general anesthesia and the shortcomings of relying solely on heart rate as an indicator of nociception.
Conclusions
This study examines the comparative effectiveness of three nociception monitoring methodsâNOL, SCA, and HRâduring standardized analgesia. NOL and SCA demonstrated superior sensitivity and responsiveness to nociceptive stimuli compared to HR, with the exception of responses during skin incision. Both NOL and SCA effectively detected significant changes in nociceptive thresholds across various stimuli, whereas HR exhibited limited variability and significance.
The findings indicate that NOL and SCA are superior to HR in detecting nociceptive changes, with NOL providing additional predictive value related to patient weight. These results are significant as they offer valuable insights for clinicians, enabling the selection of more reliable nociception monitoring techniques to enhance patient care during procedures requiring analgesia.
Limitations
We recognized that there were different personnel performing the anesthesia. However, everyone followed the departmental routine, and no deviations that could affect the results were noted. We included a variety of minimally invasive abdominal interventions, but all patients included in the study received the same type of stimulus after anesthesia induction, except for the PTC group, which underwent a tissue incision in the abdomen instead of the groin. However, their nociceptive responses did not differ from those of the other groups and did not influence the overall results.
Data availability
Data is provided within the manuscript or supplementary information files.
Abbreviations
- NOL:
-
Nociception Level Index
- SCA:
-
Skin Conductance Algesimeter
- HR:
-
Heart Rate
- ASA:
-
American Society of Anesthesiologists physical status classification system
- MAC:
-
Minimal Alveolar Concentration
- EtCO2:
-
End-Tidal Carbon Dioxide
- SBP:
-
Systolic Blood Pressure
- DBP:
-
Diastolic Blood Pressure
- MAP:
-
Mean Arterial Pressure
- SEF:
-
Spectral Edge Frequency
- SpO2:
-
Oxygen Saturation
- PRD:
-
Pupillary Reflex Dilation
- PTC:
-
Percutaneous Transhepatic Cholangiogram
- TACE:
-
Transarterial Chemoembolization
References
Ballard C, Jones E, Gauge N, Aarsland D, Nilsen OB, Saxby BK, et al. Optimised anaesthesia to reduce post operative cognitive decline (POCD) in older patients undergoing elective surgery, a randomised controlled trial. PLoS ONE. 2012;7:e37410.
Glumac S, Kardum G, Karanovic N. Post-operative cognitive decline after cardiac surgery: a narrative review of current knowledge in 2019. Med Sci Monit. 2019;25:3262â70.
World Health Organization. World report on ageing and health. Geneva: World Health Organization; 2015.
Colvin LA, Bull F, Hales TG. Perioperative opioid analgesia-when is enough too much? A review of opioid-induced tolerance and hyperalgesia. Lancet. 2019;393:1558â68.
Coccolini F, Corradi F, Sartelli M, Coimbra R, Kryvoruchko IA, Leppaniemi A, et al. Post-operative pain management in non-traumatic emergency general surgery: WSES-GAIS-SIAARTI-AAST guidelines. World J Emerg Surg. 2022;17:50.
Fletcher D, Martinez V. Opioid-induced hyperalgesia in patients after surgery: a systematic review and a meta-analysis. Br J Anaesth. 2014;112:991â1004.
Augustinsson A, Franze E, Almqvist M, WarrĂ©n Stomberg M, Sjöberg C, JildenstĂ„l P. Red-haired peopleâs altered responsiveness to pain, analgesics, and hypnotics: myth or fact?-A narrative review. J Pers Med. 2024;14:583.
Patterson Silver Wolf DAPS, Gold M. Treatment resistant opioid use disorder (TROUD): definition, rationale, and recommendations. J Neurol Sci. 2020;411:116718.
Cepeda MS, Farrar JT, Roa JH, Boston R, Meng QC, Ruiz F, et al. Ethnicity influences morphine pharmacokinetics and pharmacodynamics. Clin Pharmacol Ther. 2001;70:351â61.
Kutafina E, Becker S, Namer B. Measuring pain and nociception: through the glasses of a computational scientist. Transdisciplinary overview of methods. Front Netw Physiol. 2023;3:1099282.
Cowen R, Stasiowska MK, Laycock H, Bantel C. Assessing pain objectively: the use of physiological markers. Anaesthesia. 2015;70:828â47.
Martinez-Vazquez P, Jensen EW. Different perspectives for monitoring nociception during general anesthesia. Korean J Anesthesiol. 2022;75:112â23.
Guo J, Zhu W, Shi Q, Bao F, Xu J. Effect of surgical pleth index-guided analgesia versus conventional analgesia techniques on fentanyl consumption under multimodal analgesia in laparoscopic cholecystectomy: a prospective, randomized and controlled study. BMC Anesthesiol. 2021;21:167.
Allen J, Murray A. Age-related changes in the characteristics of the photoplethysmographic pulse shape at various body sites. Physiol Meas. 2003;24:297â307.
Storm H, Myre K, Rostrup M, Stokland O, Lien MD, Raeder JC. Skin conductance correlates with perioperative stress. Acta Anaesthesiol Scand. 2002;46:887â95.
Edry R, Recea V, Dikust Y, Sessler DI. Preliminary intraoperative validation of the nociception level index: a non-invasivee nociception monitor. Anesthesiology. 2016;125:193â203.
Bornemann-Cimenti H, Lang-Illievich K, Kovalevska K, Brenna CTA, Klivinyi C. Effect of nociception level indexâguided intraâoperative analgesia on early post-operative pain and opioid consumption: a systematic review and metaâanalysis. Anaesthesia. 2023;78:1493â501.
Munsters J, Walan JT, Holmström G, Sindelar R, Munsters, et al. Skin conductance measurement as a selective and continuous pain assessment method during eye examinations for retinopathy of prematurity. Glob Pediatr. 2023;4:100056.
Rantanen M, Yli-Hankala A, van Gils M, YppĂ€rilĂ€-Wolters H, Takala P, Huiku M, et al. Novel multiparameter approach for measurement of nociception at skin incision during general anaesthesia. Br J Anaesth. 2006;96:3:367â76.
Martini CH, Boon M, Broens SJL, Hekkelman EF, Oudhoff LA, Buddeke AW, et al. Ability of the nociception level, a multiparameter composite of autonomic signals, to detect noxious stimuli during propofol-remifentanil anesthesia. Anesthesiology. 2015;123:524â34.
Eleveld DJ, Colin P, Absalom AR, Struys MMRF. Target-controlled-infusion models for remifentanil dosing consistent with approved recommendations. Br J Anaesth. 2020;125:483â91.
JildenstÄl P, Hallén K, Almskog L, Sand J, Warren Stomberg M. Monitoring the nociception level intraoperatively-an initial experiences. J Anaesth Intensive Care Med. 2018;7:555709.
Bollag L, Jelacic S, Delgado Upegui C, Wu C, Richebe P. The nociception level index (NOL) response to intubation and incision in patients undergoing video-assisted thoracoscopic surgery (VATS) with and without thoracic epidural analgesia. A pilot study. F1000Res. 2018;7:875.
Gjerstad AC, Storm H, Hagen R, Huiku M, Qvigstad E, Raeder J. Comparison of skin conductance with entropy during intubation, tetanic stimulation and emergence from general anaesthesia. Acta Anaesthesiol Scand. 2007;51:8â15.
Park S-J, Kim BS, Jee D-L. Jaw-thrust induces sympathetic responses during induction of general anesthesia. Korean J Anesthesiol. 2013;65:127â31.
Storm H. Changes in skin conductance as a tool to monitor nociceptive stimulation and pain. Curr Opin Anaesthesiol. 2008;21:796â804.
Ledowski T, Bromilow J, Paech MJ, Storm H, Hacking R, Schug SA. Monitoring of skin conductance to assess postoperative pain intensity. BJA: Br J Anaesth. 2006;976:862â5.
Wang H, Wang Q, He Q, Li S, Zhao Y, Zuo Y. Current perioperative nociception monitoring and potential directions. Asian J Surg. 2024;47(6):2558â65.
Valdes AM, Abhishek A, Muir K, Zhang W, Maciewicz RA, Doherty M. Association of beta-blocker use with less prevalent joint pain and lower opioid requirement in people with osteoarthritis. Arthritis Care Res. 2017;69:7:1076â81.
Zhou L, Kwoh CK, Ran D, Ashbeck EL, Lo-Ciganic WH. Lack of evidence that beta blocker use reduces knee pain, areas of joint pain, or analgesic use among individuals with symptomatic knee osteoarthritis. Osteoarthr Cartil. 2020;28(1):53â61.
Khanna P, Chandralekha C, Pandey RK, Sharma A. Pain assessment in the critically ill mechanically ventilated adult patients: comparison between skin conductance algesimeter index and physiologic indicators. Saudi J Anaesth. 2018;12:204â8.
Storm H, StĂžen R, Klepstad P, Skorpen F, Qvigstad E, Raeder J. Nociceptive stimuli responses at different levels of general anaesthesia and genetic variability. Acta Anaesthesiol Scand. 2013;57:89â99.
Ma D, Ma J, Chen H, Mu D, Kong H, Yu L. Nociception monitors vs. standard practice for titration of opioid administration in general anesthesia: a meta-analysis of randomized controlled trials. Front Med. 2022;9:963185.
Sogut MS, Kalyoncu I, Karakaya MA, Manici M, Darçin K. Does Nociception Level Index-guided opioid administration reduce intraoperative opioid consumption? A systematic review and meta-analysis. Anesth Analg. 2024;139(5):978-85.
Chin S-H, Huang W-L, Akter S, Binks M. Obesity and pain: a systematic review. Int J Obes (Lond). 2020;44:969â79.
Tashani OA, Astita R, Sharp D, Johnson MI. Body mass index and distribution of body fat can influence sensory detection and pain sensitivity. Eur J Pain. 2017;21:1186â96.
Garcia MM, Corrales P, Huerta MĂ, Czachorowski MJ, LĂłpez-Miranda V, Medina-GĂłmez G, et al. Adults with excess weight or obesity, but not with overweight, report greater pain intensities than individuals with normal weight: a systematic review and meta-analysis. Front Endocrinol. 2024;15:1340465.
Marques Miranda CM, de Lima Campos M, Leite-Almeida H. Diet, body weight and pain susceptibility âa systematic review of preclinical studies. Neurobiol Pain. 2021;10:100066.
Ooba S, Ueki R, Kariya N, Tatara T, Hirose M. Mathematical evaluation of responses to surgical stimuli under general anesthesia. Sci Rep. 2020;10:15300.
Hirose M, Okutani H, Hashimoto K, Ueki R, Shimode N, Kariya N, et al. Intraoperative assessment of surgical stress response using nociception monitor under general anesthesia and post-operative complications: a narrative review. J Clin Med. 2022;11:6080.
StrĂžm C, Rasmussen LS, Steinmetz J. Practical management of anaesthesia in the elderly. Drugs Aging. 2016;33:765â77.
Foulon P, De Backer D. The hemodynamic effects of norepinephrine: far more than an increase in blood pressure!. Ann Transl Med. 2018;6(Suppl 1):S1.
Vaughan GM. In: Buijs RM, Swaab DF, editors. Handbook of clinical neurology. Vol. 117. Autonomic nervous system. Amsterdam: Elsevier; 2013.
Carella M, Beck F, Tran G, Bonhomme V, Franssen C. In response: Reporting response entropy along with state entropy would enhance differentiating nociception. Anesth Analg. 2021;133(3):798-9.
Zhou C, Huang X, Zhuo Z, Wu Q, Liu M, Li S. Effect of different anesthesia depths on perioperative heart rate variability and hemodynamics in middle-aged and elderly patients undergoing general anesthesia. BMC Anesthesiol. 2024;24(1):312.
Meijer FS, Niesters M, van Velzen M, Martini CH, Olofsen E, Edry R, et al. Does nociception monitor-guided anesthesia affect opioid consumption? A systematic review of randomized controlled trials. J Clin Monit Comput. 2020;34:629â41.
Ledowski T. Objective monitoring of nociception: a review of current commercial solutions. Br J Anaesth. 2019;123(2):e312â21.
Acknowledgements
None.
Funding
Open access funding provided by University of Gothenburg.
Regional development funds from VÀstra Götaland.
Author information
Authors and Affiliations
Contributions
Study conception by DWN and PJ. All authors were involved in study planning. DWN and PJ supervised the data collection. All authors analyzed data and drafted the manuscript. All authors revised and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
The ethics of the study were reviewed by the Swedish Ethical Review Authority, Uppsala department 2 medicine. Dnr 2021-06634-01. The study was conducted in accordance with the Declaration of Helsinki. All participants were informed about the study verbally and via written information and provided written consent to participate.
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Additional information
Publisherâs note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the articleâs Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the articleâs Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Norbeck, D.W., Lindgren, S., Wolf, A. et al. Reliability of nociceptive monitors vs. standard practice during general anesthesia: a prospective observational study. BMC Anesthesiol 25, 51 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12871-025-02923-4
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12871-025-02923-4